« Home « Chủ đề nghiên cứu hóa học

Chủ đề : nghiên cứu hóa học


Có 80+ tài liệu thuộc chủ đề "nghiên cứu hóa học"

Introduction to Modern Liquid Chromatography, Third Edition part 44

tailieu.vn

Figure 8.15 Use of seven experiments for the optimization of solvent-type selectivity in NPC.. Larger values of α for the solute to be purified are usually possible with NPC and silica columns, meaning that larger sample weights can be injected (other conditions the same). Finally, the use of TLC can be convenient for the preliminary assay of fractions from preparative...

Introduction to Modern Liquid Chromatography, Third Edition part 45

tailieu.vn

Figure 8.18 Separation of a mixture of derivatized oligosaccharides by HILIC with mobile phases of varying %-water. mobile phases are water-acetonitrile as indicated in the figure. Values of n in the figure for each peak refer to the number of saccharide units in the corresponding oligosaccharide. Although the ‘‘retention mechanism’’ for HILIC has received considerable attention in the literature, this...

Introduction to Modern Liquid Chromatography, Third Edition part 46

tailieu.vn

9.1.1 Other Reasons for the Use of Gradient Elution. Apart from the need for gradient elution in the case of wide-polarity-range samples like that of Figure 9.1, there are a number of other situations that favor or require the use of gradient elution:. High-molecular-weight compounds, such as peptides, proteins, and oligonu- cleotides, are usually poor candidates for isocratic separation, because...

Introduction to Modern Liquid Chromatography, Third Edition part 47

tailieu.vn

For gradient elution, the situation is more complicated—as values of k ∗ vary with column length and flow rate (Eq. For changes in column length L or flow rate F, a concomitant change in gradient time t G is the most convenient way of maintaining constant values of k ∗ and α. Just as a change in isocratic values of...

Introduction to Modern Liquid Chromatography, Third Edition part 48

tailieu.vn

Figure 9.11 Gradient separations of a peptide digest of recombinant human growth hormone.. (c) same as in (a), except a second gradient segment is added in order to accelerate elution of the last two peaks in the chromatogram. Thus the final gradient in Figure 9.11b is B in min.. Shortening run time is illustrated in Figure 9.11c for the sample...

Introduction to Modern Liquid Chromatography, Third Edition part 49

tailieu.vn

Figure 9.14 Plan for gradient method development.. [27]) The method-development requirements above are the same as discussed in Section 6.4 for isocratic separation, so we will not consider them further in this chapter.. the experiments used to arrive at a final separation (steps 3–8 of Table 9.2 and Fig. Ideally a full-range gradient (0–100% B) is preferred for the initial...

Introduction to Modern Liquid Chromatography, Third Edition part 50

tailieu.vn

As selectivity for a gradient method should have been optimized (step 5 of Table 9.2) prior to a change in column conditions, it is important to maintain the same values of k* (and α*) when changing column conditions and N*. This can be achieved by maintaining (t G F / L) constant (Eq. see Figure 9.6d – f. As long...

Introduction to Modern Liquid Chromatography, Third Edition part 51

tailieu.vn

Separation Conditions for Optimized Conditions in Figure 9.21a. The examples of Figure 9.21b can be more fully appreciated in terms of a relationship for gradient time. Similar plots as in Figure 9.21a,b result for the separation of higher molecular-weight samples, but with generally higher peak capacities and a need for still smaller particles.. Short run times are also needed in...

Introduction to Modern Liquid Chromatography, Third Edition part 52

tailieu.vn

9.1 or 9.26). (9.1) (k w ) value of isocratic k for. n increases with the number of polar groups in the solute molecule (Section 8.2). n increases with the number of polar groups in the solute molecule (Section 8.6). Thus, for a NPC separation with a 150 × 4.6-mm column, a flow rate of 2 mL/min, and gradient of...

Introduction to Modern Liquid Chromatography, Third Edition part 53

tailieu.vn

emphasize commercial computer-simulation software that can predict separation as a function of one or more experimental conditions, by means of experimental data from a few preliminary separations. for this application, two experimental runs are required prior to computer simulation. In the example of Figure 10.1, exper- iments using a mobile phase of 10 and 20% B (other conditions unchanged) are...

Introduction to Modern Liquid Chromatography, Third Edition part 54

tailieu.vn

Computer-Simulation Options. Peak tailing Can be simulated (Section 10.2.3.5). for maximum flexibility starting method development with gradient experiments often makes the most sense.. 10.2.3.2 Designated-Peak Selection. As an illustration of designated-peak selection, consider the example of Figure 10.4, but assume that it is required to assay only peaks 3, 8, and 10 in the presence of the remaining ‘‘interfering’’ peaks....

Introduction to Modern Liquid Chromatography, Third Edition part 55

tailieu.vn

Computer simulation can be used to check the effects of unin- tended variations in various conditions, for example:. The effect of a change in V D on a gradient separation is easily investigated by means of computer simulation. If computer simulation is used to optimize mobile-phase pH, it can also determine the robustness of the method with respect to small...

Introduction to Modern Liquid Chromatography, Third Edition part 56

tailieu.vn

Figure 11.5 Common integration errors. Rather than trying to find the peak-end point using the normal processes, the integrator will now automatically assign it to the forced time point for all chromatograms. 11.2.1.5 Additional Suggestions. The preceding discussion covers only a portion of the factors that contribute to a well-integrated chromatogram. For example, in the case of the adjustment made...

Introduction to Modern Liquid Chromatography, Third Edition part 57

tailieu.vn

(comprising all the constituents of the sample excluding the analyte. In applications where occasional over-limit samples are likely to be encountered, such as preliminary pharmacokinetic studies, it is wise to include dilution tests as part of the validation process. In other cases, concentration of the sample during sample pretreatment, or less dilution, can provide a solution. In any event, the...

Introduction to Modern Liquid Chromatography, Third Edition part 58

tailieu.vn

11.4.1.4 Standard Addition. can be useful when a sample blank cannot be obtained, and the sample matrix can affect analyte recovery and/or response. For example, when measuring insulin levels in plasma, it is impossible to obtain plasma without insulin, so standard addition can be used.. Standard addition can be based on a single-point or multiple-point calibration.. One fraction is spiked...

Introduction to Modern Liquid Chromatography, Third Edition part 59

tailieu.vn

Table 12.1. Table 12.1 illustrates a representative accuracy study. The data should be reported as the percent recovery of the known or added amount, or as the difference between the mean and true value with confidence intervals. In Table 12.1, data are shown relative to 100%, and the mean recovery for n = 9 samples is 98.62%. 12.2.2 Precision. 12.2.2.1...

Introduction to Modern Liquid Chromatography, Third Edition part 60

tailieu.vn

Each subsection should include a brief summary of the applicable protocol, and the mean, standard deviation, relative standard deviation, acceptance criteria, and assessment (pass or fail).. A properly designed validation protocol can serve as a template for the validation report. For example, in the protocol a test can be described and the acceptance criteria listed. For the validation report, this...

Introduction to Modern Liquid Chromatography, Third Edition part 61

tailieu.vn

12.7.1.2 Co-validation between Laboratories. Comparative testing (Section 12.7.1.1) traditionally requires a validated method as a prerequisite to AMT. However, another option for AMT is to involve the receiving laboratory from the beginning in the actual validation of the test method to be transferred. 12.7.1.3 Method Validation and/or Revalidation. Another technique that can be used for AMT involves the receiving laboratory’s...

Introduction to Modern Liquid Chromatography, Third Edition part 62

tailieu.vn

Guideline for Submitting Samples and Analytical Data for Methods Validation, USFDA-CDER (February 1987), http://www.fda.gov/cder/guidance/ameth.htm.. 2000), http://www.fda.gov/cder/guidance/2396dft.htm.. 2005), http://www.ich.org/LOB/media/MEDIA417.pdf.. Guidance for Methods Development and Methods Validation for the Resource Con- servation and Recovery Act (RCRA) Program, US EPA, (1995), http://www.epa.gov/. ISO/IEC 17025, General Requirements for the Competence of Testing and Calibration Laboratories, (2005), http://www.iso.org/iso/iso catalogue/catalogue tc/. Current Good Manufacturing Practice in...

Introduction to Modern Liquid Chromatography, Third Edition part 63

tailieu.vn

Figure 13.4 Schematic diagram of the helical structure of DNA, showing hydrogen bonding between complementary base pairs.. Exposure of the helix to elevated temperature or extremes of pH can induce separation of the two strands.. these structures can be dissociated by heat or chemically denaturing conditions.. 13.2.3 Carbohydrates. 13.5) are glucose, galactose, mannose, fucose, N-acetylgalactosamine (GalNAc), and N-acetylglucosamine (GlcNAc). Figure...